0

Anaesthesia Workstation Machine

Anaesthesia Workstation Machine
Product Description

 

 

 

SYSTEMA- 15

ANAESTHESIA MACHINE


KEY FEATURES:         

  3 gas system Hypoxic Guard-Oxygen, Air, Nitrous

  Ventilator with Dual Mode –Adult & Paediatric

  Monitor with 5 multipara facilities

  Imported ISO Fluorine/ Savo Fluorine/ Halothane Vaporizer

  Circle Absorber System

  Extra Space with dual drawers soundless ball channel systems.

 


TECHNICAL SPECIFICATIONS: 

Gas Rotameters :

MNLCP SYSTEMA 15 has 3 gases delivery system. Oxygen, Air, Nitrous. 3 Calibrated Rotameters (O2, Air, N2). It has Hypoxic guard system which deliver gases as per 3:1 Oxygen & Nitrous ratio.

 

Ventilator :    

Patient Type                        Adult and Pediatric        

Tidal Volume                       50 to 1200 ml.                

Frequency                           8 to 40 BPM  

I : E Ratio                              1:1, 1:2, 1:3 and 1:4      

Mode of Operation           CMV, Manual

Alarms                                  Pressure Low and High, Active patient, driving gas failure , low battery indication, circuit disconnect,

                                              Minute Volume low and High        

Monitoring Parameters -Tidal Volume and Minute Volume 

Pressure Display                         -10 to 90 mBar in LED graph

High Pressure Limit            20 to 60 mBar                

Low Pressure Limit             0 to 20 mBar

Power Input                        90 to 260 VAC 50/60 Hz               

Minute Volume High         3 to 50 Liters

Minute Volume Low          0 to 47 Liters

Driving Gas Input                3 Bar Pressure               

Battery Backup                    up to 2 Hours ( Lithium-ion )    

 

Multipara Monitor :

6 Standard Parameters: ECG, RESP, NIBP, SPO2, 2-Temp, PR/HR.

8” high resolution colour TFT LCP display with touch screen.

Real time S-T segment analysis, pace maker detection.

Multi lead ECG wave forms display in phase

Up to 2 hours working capacity of built in rechargeable battery

Networking capacity

Patients info input management function

Capture dynamic waveforms.

Efficient resistance to interference of defibrillator and HF knife.

Large volume of tabular and graphic trends information storage and easy to recall.

 

 

Vaporizer :     

5 Lpm O2 @ 210C ± 20C

Vaporiser 5% ±0.25 of the delivered agent or ± 15% of the dial setting (whichever is greater)

Vaporiser 8% ±0.4 of the delivered agent or ± 15% of the dial setting (whichever is greater)

Calibrated at sea level

 

Environmental Specification -

Temperature 150C to 350C
Humidity 30% to 75% relative humidity.

Works consistently in altitudes upto 2440m above sea level

Performance Reliability -

Performs with gas flow rates from 0.2 - 15 lpm

Systems are Selectatec® Compatible With Interlock & Cage mount (ISO 23mm tapers)

Types of Vaporizers - Isoflurane, Sevoflurane, Halothane, Enflurane. Can be fitted well.

 

Warranty: 1 Year

Contact Us

43, Joy Kulla Molla Road, Near Barisha Nutan Sangha Club, Kolkata, West Bengal, 700063, India
Phone :+918068211984